Temperature, humidity, and pressure converter with relay outputs, Ethernet interface, and discrete inputs. External probes can be connected. Designed for installation in a 19” 1U rack. — H7531R
Available on request
Delivery across Russia
Order phone: +7(800)777-18-50
Description
The converter is equipped with two relay outputs for signaling, indication, or control of external devices. It also has three discrete inputs for connecting external devices such as smoke detectors, motion sensors, etc. Delay, hysteresis, and sound signal settings are included for each relay. It is installed in a 19” 1U rack. Parameters are easily configured using a computer or the keyboard on the sensor's front panel. The converter is controlled by a microprocessor housed in a durable plastic case with terminal connectors. A large two-line LCD display is designed to simultaneously show two parameters. The display can be turned off if necessary.
Recently, temperature and humidity levels have been considered as the most important parameters responsible for indoor climate quality. Especially in buildings and rooms with large gatherings of people, such as conference halls, hospitals, schools, cinemas, theaters, and wellness centers.
With the help of such sensors, engineers, environmental experts, and healthcare specialists can optimize ventilation to create a comfortable and "Healthy" indoor climate.
ATTENTION, the device is currently discontinued.
- Specifications
- Overall Drawing
- Electrical diagram
- Applications
|
Sensor |
H7531R |
|
|
Output signal |
2x Relay + Ethernet |
|
|
Temperature |
Range |
-30 to +105°C |
|
Accuracy |
±0.4°C |
|
|
Relative Humidity (at RH 5-95% and 23°C) |
Range |
0 to 100 %RH |
|
Accuracy |
±2.5 %RH |
|
|
Atmospheric Pressure |
Range |
600 to 1100 hPa |
|
Accuracy |
±1.3 hPa |
|
|
Discrete input |
yes |
|
|
Maximum switchable voltage, current and power by relay output |
50V, 2A, 60VA |
|
|
Accuracy and range of dew point temperature measurement |
±1.5°C at ambient temp. T<25°C and RH >30% |
|
|
Accuracy and range of absolute humidity measurement |
±1.5 g/m3 at ambient temp. <250C, range 0 to 400 g/m3 |
|
|
Accuracy and range of specific humidity measurement |
±2.1 g/kg at ambient temp. T < 35°C |
|
|
Accuracy and range of specific enthalpy measurement |
±3 kJ/kg at ambient temp. T < 25°C |
|
|
Accuracy and range of mixture component ratio measurement |
±2 g/kg at ambient temp. T < 35°C |
|
|
Power supply voltage |
9 - 30 V |
|
|
Sensor / external sensor protection class |
IP30/IP40 |
|
|
Operating temperature range of sensor |
From -30 to +80°C |
|
|
Operating temperature range of external sensor |
From -30 to +105°C |
|
|
Operating humidity range (without condensation) |
0 to 100 %RH |
|
|
Operating atmospheric pressure range |
600 to 1100 hPa |
|
|
Storage temperature range |
From -30 to +80°C |
|
|
Recommended calibration interval |
1 year |
|
|
Dimensions / Weight |
483 x 44 x 45 mm (W x H x D), probe length 88 mm, diameter 18 mm / 410 g |
|
|
Warranty |
3 years |
|




Power supply using PoE splitter

Connecting relay to sensor outputs

Relay output parameters
Voltage – max 50 V
Power – max 60 VA
Current – max 2 A