Temperature, humidity, and CO2 concentration converters with RS232, RS485 interfaces and two relay outputs. — Sensors H6320 and H6420

Available on request
Delivery across Russia
Order phone: +7(800)777-18-50

0 RUB

Description

Temperature, humidity, and CO2 concentration converters with RS232, RS485 interfaces. Equipped with two relay outputs for signaling, indication, or control of external devices. Delay, hysteresis, and sound signal settings are included for each relay. Parameters are easily adjustable using a computer or the control keyboard. The converter is microprocessor-controlled, housed in a durable plastic case with terminal connections. A large two-line LCD display is designed for simultaneous display of two parameters. The display can be turned off if necessary.

Recently, levels of carbon dioxide, temperature, and humidity are considered key parameters responsible for indoor climate quality. Especially in buildings and rooms with large gatherings of people, such as conference halls, hospitals, schools, cinemas, theaters, and wellness centers.

With the help of such sensors, engineers, environmental experts, and healthcare specialists can optimize ventilation to create a comfortable and "Healthy" indoor climate.

ATTENTION, the device is currently discontinued.

  • Specifications
  • Overall Drawing
  • Electrical diagram
  • Applications

Sensors

H6320

H6420

Output

RS232

RS485

Measured values

CO2 + Relative Humidity + Temperature

Measured temperature range

from -30 to +60°C

Relay output

Yes

Discrete input

No

LCD display

Yes

PoE

No

Max. switchable voltage, current and power by relay output

50V, 2A, 60VA

CO2 measurement range

from 0 to 2000ppm (optional 5000 ppm, 10000 ppm)

CO2 concentration measurement accuracy

±(50ppm +2% of measured value) at 25°C and 1013hPa

Relative humidity measurement range

from 0 to 100%

Relative humidity measurement accuracy

±2.5% at relative humidity from 5 to 95% and at 23°C

Temperature measurement accuracy

±0.4°C

Available temperature units

Celsius, Fahrenheit

Dew point temperature measurement accuracy and range

±1.5°C at ambient temp. T<25°C and RH>30%
range from -60 to +80 °C

Absolute humidity measurement accuracy and range

±1.5g/m3 at ambient temp. T < 25°C
range from 0 to 400 g/m3

Specific humidity measurement accuracy and range

±2g/kg at ambient temp. T < 35°C
range from 0 to 550 g/kg

Mixture component ratio measurement accuracy and range

±2g/kg at ambient temp. T < 35°C
range from 0 to 995 g/kg

Specific enthalpy measurement accuracy and range

±3kJ/kg at ambient temp. T < 25°C
range from 0 to 995 kJ/kg

CO2 measurement interval

15 sec

Operating temperature range

from -30 to +60°C

Filter mesh capability on sensor housing

0.025mm – stainless steel mesh

Protocol

ModBus RTU and Advantech ADAM compatible protocol

Sensor protection class

IP30 / IP40

Power supply voltage

9-30 V DC, power consumption approximately 1W

Dimensions

136 x 213 x 45 mm (W x H x D), rod length 75 mm

Weight

360 g

Warranty

3 years

 