Russian manufacturer
of instrumentation and control equipment
EN

8 (800) 777 18 50

Working hours

  • Monday
    09:00-18:00
  • Tuesday
    09:00-18:00
  • Wednesday
    09:00-18:00
  • Thursday
    09:00-18:00
  • Friday
    09:00-18:00
  • Saturday
    Closed
  • Sunday
    Closed
Survey sheets
Request a quote
  • Monday
    09:00-18:00
  • Tuesday
    09:00-18:00
  • Wednesday
    09:00-18:00
  • Thursday
    09:00-18:00
  • Friday
    09:00-18:00
  • Saturday
    Closed
  • Sunday
    Closed

8 (800) 777 18 50

EN
Request a quote

Temperature, humidity, and CO2 concentration converters with RS232, RS485 interfaces, two relay outputs, and remote probe. — Sensors H6321 and H6421

Available on request
Delivery across Russia
Order phone: +7(800)777-18-50

0 RUB

Description

Temperature, humidity, and CO2 concentration converters with RS232, RS485 interfaces and remote probe with various cable lengths (1, 2, 4m). Equipped with two relay outputs for signaling, indication, or control of external devices. Delay, hysteresis, and sound signal settings are included for each relay. Parameters are easily adjustable using a computer or with the control keyboard. The converter is microprocessor-controlled, housed in a durable plastic case with terminal connectors. A large two-line LCD display is designed for simultaneous display of two parameters. The display can be turned off if necessary.

Recently, levels of carbon dioxide, temperature, and humidity are considered as the most important parameters responsible for indoor climate quality. Especially in buildings and rooms with large gatherings of people, such as conference halls, hospitals, schools, cinemas, theaters, and wellness centers.

With the help of such sensors, engineers, environmental experts, and healthcare specialists can optimize ventilation to create a comfortable and "Healthy" indoor climate.

ATTENTION, the device is currently discontinued.

  • Specifications
  • Overall Drawing
  • Electrical diagram
  • Applications

Sensors

H6321

H6421

Outputs

RS232

RS485

Measured values

CO2 + Relative Humidity + Temperature

Measured temperature range

from -30 to +105°C

Relay output

Yes

Discrete input

No

LCD display

Yes

PoE

No

Max. switchable voltage, current and power by relay output

50V, 2A, 60VA

CO2 measurement range

from 0 to 10000ppm (optional 2000 ppm, 5000 ppm)

CO2 concentration measurement accuracy

±100ppm +5% of measured value at 25°C and 1013hPa

Relative humidity measurement range

from 0 to 100%

Relative humidity measurement accuracy

±2.5% at relative humidity from 5 to 95% and at 23°C

Temperature measurement accuracy

±0.4°C

Available temperature units

Celsius, Fahrenheit

Dew point temperature measurement accuracy and range

±1.5°C at ambient temp. T<25°C and RH>30%
range from -60 to +80 °C

Absolute humidity measurement accuracy and range

±1.5g/m3 at ambient temp. T < 25°C
range from 0 to 400 g/m3

Specific humidity measurement accuracy and range

±2g/kg at ambient temp. T < 35°C
range from 0 to 550 g/kg

Mixture component ratio measurement accuracy and range

±2g/kg at ambient temp. T < 35°C
range from 0 to 995 g/kg

Specific enthalpy measurement accuracy and range

±3kJ/kg at ambient temp. T < 25°C
range from 0 to 995 kJ/kg

CO2 measurement interval

15 sec

Operating temperature range CO2 probe / RH+T probe

from -30 to +60°C / from -30 to +105°C

Filtering capability of sensor housing mesh

0.025mm – stainless steel mesh

Protocol

ModBus RTU and Advantech ADAM compatible protocol

Protection class sensor / CO2 probe / RH+T probe

IP65 / IP65 / IP40

Power supply voltage

9-30 V DC, power consumption approximately 1W

Dimensions

136 x 136 x 45 mm (W x H x D)

Weight without probe / with 1m / 2m / 4m probe

360 / 520 / 590 / 730g

Warranty

3 years

 

 

Connecting the relay to the sensor outputs

 

Send request

Request a callback

Information sent successfully

Благодарим за обращение.
Ваша заявка будет рассмотрена в рабочее время:
Пн-Пт с 9 до 18 по московскому времени.