Ethylene Concentration Converter — E2618-C2H4
Available on request
Delivery across Russia
Order phone: +7(800)777-18-50
Description
Ethylene Concentration Transmitter E2618-C2H4— a representative of the new line of multifunctional measuring instruments PluraSens®.
The transmitter can be used in the polymer industry, agriculture, and other fields where accurate measurement of ethylene concentration is required.
The device is equipped with a fully calibrated latest-generation electrochemical sensor with temperature compensation, ensuring high stability and reliability of readings, as well as a long service life. Thanks to humidity compensation (optional), the device can be used at relative humidity up to 100%.
Two analog outputs and an RS485 interface operating with the Modbus RTU protocol are used to connect the transmitter.
Advantages
- Accurate and stable measurement
- Wall-mounted or duct version
- Remote probe (optional)
- 2 configurable analog outputs 4-20 mA or 0-10 V, Digital RS485 Modbus RTU output
- Display version available
ATTENTION, the device is currently discontinued.
- Specifications
- Overall dimensions drawing
- Connection diagram
- Applications
| Specifications | |
| Calibration | Ethylene C2H4 |
| Sensor type | Electrochemical |
| Detection method | Diffusion |
| Measurement range |
0...10 ppm C2H4 (peak 20 ppm) 0...200 ppm C2H4 (peak 500 ppm) 0...1500 ppm C2H4 (peak 2000 ppm) |
| Resolution | 0.1 ppm C2H4 1 ppm C2H4 1 ppm C2H4 |
| Repeatability | <2 % of signal |
| Response time T80 | < 60 seconds |
| Service interval | 6 months |
| Sensor lifetime | 2 years |
| Self-diagnostics | Full functionality check at startup |
| Warm-up time | ≤ 1 min |
| Power supply | 11...30 V DC |
| Power consumption | < 2 VA |
| Digital interface | RS485, Modbus RTU protocol |
| Analog outputs | 2 × 4-20 mA / 0-10 V, configurable |
| Housing | Light gray ABS plastic, IP65 protection rating |
| Dimensions | 82 × 80 × 55 mm |
| Remote probe | Protection rating IP20, Cable length standard 3.0 m |
| Applications | Industrial premises, outdoor use |
| Operating conditions | 20..+50 °C, 15...90 %RH; Atmospheric pressure ±10%; not for explosive zones; no strong mechanical impacts, vibrations or EMI Non-aggressive environment |
| Display (optional) | |
| Operating temperature | 0...+50 °C |
| Display size | 72 × 36 mm |
| Number of characters | 3.5 7-segment |
| Character height | 14 mm |
| Other features | Backlight |


You may find useful