Ethylene Concentration Converter — E2618-C2H4

Available on request
Delivery across Russia
Order phone: +7(800)777-18-50

0 RUB

Description

Ethylene Concentration Transmitter E2618-C2H4— a representative of the new line of multifunctional measuring instruments PluraSens®.

The transmitter can be used in the polymer industry, agriculture, and other fields where accurate measurement of ethylene concentration is required. 

The device is equipped with a fully calibrated latest-generation electrochemical sensor with temperature compensation, ensuring high stability and reliability of readings, as well as a long service life. Thanks to humidity compensation (optional), the device can be used at relative humidity up to 100%.

Two analog outputs and an RS485 interface operating with the Modbus RTU protocol are used to connect the transmitter.

Advantages

  • Accurate and stable measurement
  • Wall-mounted or duct version
  • Remote probe (optional)
  • 2 configurable analog outputs 4-20 mA or 0-10 V, Digital RS485 Modbus RTU output
  • Display version available

ATTENTION, the device is currently discontinued.

  • Specifications
  • Overall dimensions drawing
  • Connection diagram
  • Applications
Specifications
Calibration Ethylene C2H4
Sensor type Electrochemical
Detection method Diffusion
Measurement range

0...10 ppm C2H4 (peak 20 ppm)

0...200 ppm C2H4 (peak 500 ppm)

0...1500 ppm C2H4 (peak 2000 ppm)

Resolution 0.1 ppm C2H4
1 ppm C2H4
1 ppm C2H4
Repeatability <2 % of signal
Response time T80 < 60 seconds
Service interval 6 months
Sensor lifetime 2 years
Self-diagnostics Full functionality check at startup
Warm-up time ≤ 1 min
Power supply 11...30 V DC
Power consumption < 2 VA
Digital interface RS485, Modbus RTU protocol
Analog outputs 2 × 4-20 mA / 0-10 V, configurable
Housing Light gray ABS plastic, IP65 protection rating
Dimensions 82 × 80 × 55 mm
Remote probe Protection rating IP20,
Cable length standard 3.0 m
Applications Industrial premises, outdoor use
Operating conditions 20..+50 °C, 15...90 %RH;
Atmospheric pressure ±10%; not for
explosive zones; no strong mechanical impacts, vibrations or EMI
Non-aggressive environment
Display (optional)
Operating temperature 0...+50 °C
Display size 72 × 36 mm
Number of characters 3.5 7-segment
Character height 14 mm
Other features Backlight