Industrial pressure transmitter, 0.25%FS — 23SY
Available on request
Delivery across Russia
Order phone: +7(800)777-18-50
Description
Pressure transmitter 23SY – an industrial pressure transmitter that uses the proven and stable Keller sensor series 9L, with electronics enhanced for electromagnetic field resistance. According to international CE standards, the sensors can be used for field levels of 10.
Thanks to an additional electrical circuit, this sensor has very low temperature error, and the fully welded design eliminates the need for internal seals. Thus, we have a reliable sensor for critical applications, capable of withstanding vibrations up to 20g (5…2000Hz). Ideal for reliable and stable measurement in liquid and gaseous media, and also, thanks to the possibility of using an IP68 cable output, this sensor can be used as a hydrostatic level meter.
Options
- Threads ¼”NPT, G1/2”, G1/2” Mano
- Any other ranges within -1…1000 bar
- Filling with fluorocarbon oil (O2-compatible), low temperature oil (-55 °C)
- Other connectors for connection, IP68 cable output
- Specifications
- Overall Drawing
- Connection diagram
- Applications
|
Specification |
Value |
|
Pressure range (all ranges within specified are available) * |
PR: from -0.1…0.1 bar to 0…30 bar |
|
Output signal |
4…20 mA / 0…10 V |
|
Basic accuracy |
0.25% FSO |
|
Operating temperature |
-40…120°C |
|
Connection |
G1/4”, 1/4”-18NPT, G1/2”, G1/2” Mano |
|
Electrical connection |
Connectors: Binder-581/680 (5-pin, IP 40),Binder-423/723 (5-pin, IP67), GSP (DIN 43650), M12 (4-pin), Souriau (MIL-C26482) |
|
Material |
Stainless steel 316L |
|
Overloads |
x2 measurement range |
|
Bandwidth |
2 kHz |
|
Stability, typ. |
0.1% FSO |
* - any special ranges are available, for example: 0.5…10 bar or -1…24 bar
PR – is a relative pressure transmitter, zero reading at atmospheric pressure. Can be used for measuring both gauge pressure and vacuum pressure.
PAA – is an absolute pressure transmitter.
PA – is a gauge pressure transmitter (zero at 1 bar abs.). Suitable for use where atmospheric pressure changes can be neglected (for ranges above 10 bar) or where a hermetic sensor is required.
|
|
|
Total error |
Total error |
||
|
Measurement ranges, FSO |
Basic accuracy |
Typical |
Max. |
Typical |
Max. |
|
2…1000 bar |
0.25 % FSO |
±0.3% FSO |
±0.5% FSO |
±0.4 % FSO |
±0.7% FSO |
|
0.5…< 2 bar |
±0.25 %FSO |
±0.6 %FSO |
±1.0 %FSO |
±0.8 %FSO |
±1.5 %FSO |
|
0.1…0.5 bar |
±0.50 %FSO |
± 2 mbar |
± 3 mbar |
± 2.5 mbar |
± 4.5 mbar |


|
|
2-wire |
3-wire, |
|
Supply voltage |
8…32 VDC |
13…32 VDC |
|
Output signal |
4…20mA |
0…10V / 0.5…4.5 V / 0…5V |
|
Load resistance |
RL < (Ui-8) / 0.025 A |
RL > 5 kΩ |
|
Current consumption |
max. 25 mA |
max. 4 mA |